Assessment of shoulder range of motion using a wireless inertial motion capture device - a validation study

Download (415.07 kB)
journal contribution
posted on 2019-04-13, 00:00 authored by Michael Rigoni, Stephen GillStephen Gill, Sina Babazadeh, Osama Elsewaisy, Hugh Gillies, Nhan Nguyen, Pubudu PathiranaPubudu Pathirana, Richard PageRichard Page
(1) Background: Measuring joint range of motion has traditionally occurred with a universal goniometer or expensive laboratory based kinematic analysis systems. Technological advances in wearable inertial measurement units (IMU) enables limb motion to be measured with a small portable electronic device. This paper aims to validate an IMU, the 'Biokin', for measuring shoulder range of motion in healthy adults; (2) Methods: Thirty participants completed four shoulder movements (forward flexion, abduction, and internal and external rotation) on each shoulder. Each movement was assessed with a goniometer and the IMU by two testers independently. The extent of agreement between each tester's goniometer and IMU measurements was assessed with intra-class correlation coefficients (ICC) and Bland-Altman 95% limits of agreement (LOA). Secondary analysis compared agreement between tester's goniometer or IMU measurements (inter-rater reliability) using ICC's and LOA; (3) Results: Goniometer and IMU measurements for all movements showed high levels of agreement when taken by the same tester; ICCs > 0.90 and LOAs < ±5 degrees. Inter-rater reliability was lower; ICCs ranged between 0.71 to 0.89 and LOAs were outside a prior defined acceptable LOAs (i.e., > ±5 degrees); (4) Conclusions: The current study provides preliminary evidence of the concurrent validity of the Biokin IMU for assessing shoulder movements, but only when a single tester took measurements. Further testing of the Biokin's psychometric properties is required before it can be confidently used in routine clinical practice and research settings.

History

Journal

Sensors

Volume

19

Issue

8

Article number

1781

Publisher

MDPI

Location

Basel, Switzerland

eISSN

1424-8220

Language

eng

Publication classification

C1 Refereed article in a scholarly journal

Copyright notice

2019, The Authors