posted on 2006-01-01, 00:00 authored by W Indrasena, Colin BarrowColin Barrow, J Kralovec
The combined effect of scan speed, hydrogen and air flow rates on the flame ionization detection (FID) peak response of phospholipid classes has been studied to determine the optimum levels of these parameters. The phospholipid composition of different types of commercial lecithins, as well as lecithins combined with fish oils, has been analyzed by Iatroscan TLC‐FID Mark‐6s under optimized conditions. An air flow rate of 2 L/min, a hydrogen flow rate of 150–160 mL/min, and a scan speed of 30 s/rod seem to be the ideal conditions for scanning phospholipids with complete pyrolysis in the flame in the Mark‐6 model. Increasing the scan speed rapidly decreased the FID response. A hydrogen flow rate as high as 170 mL/min could be used at relatively low air flow rates (ς L/min) and the response declined when both air flow rate and hydrogen flow rate increased simultaneously. Both linear and curvilinear relationships had highly significant correlations (pπ.01) with the sample load. Time course reactions, including the hydrolysis of phosphatidylserine using enzymes, can be successfully monitored by the Iatroscan TLC‐FID Chromarod system.

History

Journal

Journal of liquid chromatography and related technologies

Volume

29

Issue

14

Pagination

2111 - 2127

Publisher

Marcel Dekker

Location

Monticello, N.Y.

ISSN

1082-6076

eISSN

1520-572X

Language

eng

Publication classification

C1.1 Refereed article in a scholarly journal

Copyright notice

2006, Taylor & Francis Group

Usage metrics

    Research Publications

    Categories

    No categories selected

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC