Measurement of immunoglobulin A in saliva by particle-enhanced nephelometric immunoassay: sample collection, limits of quantitation, precision, stability and reference range
journal contribution
posted on 2009-09-01, 00:00 authored by C Booth, Dan DwyerDan Dwyer, P Pacque, M BallBackground: Total immunoglobulin A in saliva (s-IgA) is normally assayed using an enzyme-linked immunosorbent assay. We have investigated methodological issues relating to the use of particle-enhanced nephelometric immunoassay (PENIA)
to measure s-IgA in whole unstimulated saliva and determine its reference range.
Methods: Whole unstimulated resting saliva was collected to determine sample stability (temperature, time, effect of a protease inhibitor), limit of quantitation (LOQ), assay precision and analytical variation. The reference range for 134 healthy adults was determined.
Results: Linearity was excellent (4–10.3 mg L21, P, 0.001; R2 ¼ 0.997) and without significant bias (mean of 20.7%). The lowest intra- and inter-analytical coefficients of variation were 1.8% and 7.5% and LOQ was 1.4 mg L21. The concentration of s-IgA is stable at room temperature for up to 6 h, at 48C for 48 h, at 248C for two weeks and at 2808C for up to 1.3 yr. There is no evidence that a protease inhibitor increases the stability or that repeated freeze–thawing cycles degrade sample quality. The reference ranges for s-IgA concentration, s-IgA secretion, s-IgA:albumin and s-IgA:osmolality were 15.9–414.5 mg L21, 7.2–234.9 mg min21, 0.4–19 and 0.6–8.9, respectively.
Conclusion: Automated PENIA assay of s-IgA is precise and accurate. High stability of collected saliva samples and the ease and speed of the assay make this an ideal method for use in athletic and military training situations. The convenience of measuring albumin and IgA on the same analytical platform adds to the practicability of the test.
to measure s-IgA in whole unstimulated saliva and determine its reference range.
Methods: Whole unstimulated resting saliva was collected to determine sample stability (temperature, time, effect of a protease inhibitor), limit of quantitation (LOQ), assay precision and analytical variation. The reference range for 134 healthy adults was determined.
Results: Linearity was excellent (4–10.3 mg L21, P, 0.001; R2 ¼ 0.997) and without significant bias (mean of 20.7%). The lowest intra- and inter-analytical coefficients of variation were 1.8% and 7.5% and LOQ was 1.4 mg L21. The concentration of s-IgA is stable at room temperature for up to 6 h, at 48C for 48 h, at 248C for two weeks and at 2808C for up to 1.3 yr. There is no evidence that a protease inhibitor increases the stability or that repeated freeze–thawing cycles degrade sample quality. The reference ranges for s-IgA concentration, s-IgA secretion, s-IgA:albumin and s-IgA:osmolality were 15.9–414.5 mg L21, 7.2–234.9 mg min21, 0.4–19 and 0.6–8.9, respectively.
Conclusion: Automated PENIA assay of s-IgA is precise and accurate. High stability of collected saliva samples and the ease and speed of the assay make this an ideal method for use in athletic and military training situations. The convenience of measuring albumin and IgA on the same analytical platform adds to the practicability of the test.
History
Journal
Annals of clinical biochemistryVolume
46Issue
5Pagination
401 - 406Publisher
SageLocation
London, EnglandPublisher DOI
Link to full text
ISSN
0004-5632Language
engPublication classification
C1.1 Refereed article in a scholarly journalCopyright notice
2009, Sage PublicationsUsage metrics
Categories
No categories selectedKeywords
Licence
Exports
RefWorks
BibTeX
Ref. manager
Endnote
DataCite
NLM
DC