Multiplexed detection: fast comprehensive sample analysis of tobacco leaf extracts using HPLC with AFT columns

Version 2 2024-06-03, 12:24
Version 1 2016-01-07, 11:31
journal contribution
posted on 2024-06-03, 12:24 authored by S Pravadali-Cekic, D Kocic, Xavier ConlanXavier Conlan, RA Shalliker
Detector-based comprehensive screening analysis of complex samples of natural origin using High Performance Liquid Chromatography (HPLC) can be a complicated and time-consuming task. There are a number of ways multidetection characterization can be achieved; however, there are limitations associated with each technique. Active Flow Technology (AFT) in Parallel Segmented Flow (PSF) mode allows for multiplexed detection HPLC analysis within a single injection, whereas maintaining chromatographic performance and allowing the use of multiple destructive detectors to achieve a comprehensive yet efficient screening of a complex sample. In this study, a comprehensive characterization analysis of tobacco leaf extract was carried out through multiplexed detection using a PSF column for the detection of biomolecules by UV-Vis detection, DPPH• for reactive-oxygen species (ROS) detection, and mass spectrometry, the latter two detection methods being sample destructive.

History

Journal

Journal of liquid chromatography and related technologies

Volume

38

Pagination

1753-1758

ISSN

1082-6076

eISSN

1520-572X

Language

eng

Publication classification

C Journal article, C1 Refereed article in a scholarly journal

Copyright notice

2015, Taylor & Francis

Issue

19

Publisher

Taylor & Francis