The effect of three different miniaturized blood purification devices on plasma cytokine concentration in an ex vivo model of endotoxinemia

Version 2 2024-06-03, 23:34
Version 1 2008-01-01, 00:00
journal contribution
posted on 2024-06-03, 23:34 authored by M Haase, R Bellomo, Ian Baldwin, A Haase-Fielitz, M Storr, B Krause, N Boyce, S Svobodova, W Li, SM Bagshaw, S Warrillow, C Langenberg, S Morgera
Purpose A novel type of adsorptive plasma filtering device (ETX-A) capable of removing endotoxin from blood in a single step has recently been developed using nanotechnology. Methods In a miniaturized, ex vivo model of extracorporeal circuits, we tested the capacity to reduce plasma cytokine concentration of ETX-A filters in comparison to standard high-flux (HF) filters, high cut-off (HCO) filters and a control. Blood from six healthy volunteers was spiked with endotoxin and then circulated through closed (ETX-A, control) or open (HF, HCO) circuits. Blood flow was set at 16 ml/min and filtration flow at 1 ml/min. Samples for measurement of IL-1ra and IL-6 were taken at baseline and at 4 hours. Results Compared to control (703.3 [850.6] pg/mL), in HCO (383.5 [1144.1] pg/mL) and ETX-A (490.1 [683.2] pg/mL) filters, plasma IL-1ra pooled pre- and postfilter concentrations were lower at the end of the experiment (P=0.002; P=0.050, respectively) whereas, in standard HF filters, IL-1ra concentration was higher than control. HCO showed a trend toward a reduced relative increase in IL-6 concentration from commencement to end of experiment compared to control (P=0.07). After pooling end-of-experiment plasma cytokine values of novel blood purification devices, we found HCO + ETX-A superior to H with regard to reduction of IL-1ra (-27.0 [−20.5]% vs. 8.1 [18.9]%; P<01) and IL-6 (-18.0 [38.3]% vs. −1.1 [24.3]%; P=0.050) compared to control. Conclusions HCO and ETX-A appeared to significantly reduce plasma IL-1ra and, when combined, plasma IL-6 concentration as well. It appears desirable to manufacture full-size blood purification devices using this technology and to explore their effect on cytokine removal.

History

Related Materials

Location

United States

Language

en

Publication classification

C1.1 Refereed article in a scholarly journal

Journal

International Journal of Artificial Organs

Volume

31

Pagination

722-729

ISSN

0391-3988

eISSN

1724-6040

Issue

8

Publisher

SAGE Publications