Deakin University
Browse

The streak retinoscopy pupil reflex in the presence of astigmatism

journal contribution
posted on 2003-07-01, 00:00 authored by G Smith, S Haymes
Computer simulation is being increasingly used as a teaching tool. Having developed a computer-generated virtual focimeter, we are now in the process of developing a computer-generated virtual streak retinoscope to teach the principles of retinoscopy and the effect of residual refractive error and mirror movement on the pupil reflex. One of the important requirements was to provide as accurate a simulation as possible for the completely general case of an astigmatic patient and a streak in any orientation being moved also in any orientation. This required a thorough understanding of the optical theory of the retinoscope and equations that describe the behaviour of the pupil reflex. We have taken this opportunity to review the optics of the streak retinoscope and derive equations for the behaviour of the pupil streak reflex.

History

Journal

Ophthalmic and physiological optics

Volume

23

Issue

4

Pagination

295 - 305

Publisher

Wiley - Blackwell Publishing

Location

Oxford, England

ISSN

0275-5408

eISSN

1475-1313

Language

eng

Publication classification

C1.1 Refereed article in a scholarly journal

Copyright notice

2003, The College of Optometrists 295